Insulation work of medical device shell

Release time:2023-03-30

Medical diagnosis, measurement and treatment equipment using AC power supply may cause leakage current due to improper grounding and electrical insulation, which may expose patients and even medical staff to the risk of electric shock, burns, internal organ damage and arrhythmia. So do a good jobMedical device shellInsulation work.

美容儀器殼9.jpg


1. The equipment shall be manufactured and sealed to prevent contact with live parts and parts that may be live in a single fault state. If the bulb can be replaced without tools, it must be ensured to prevent contact with the live parts of the lamp when loading and unloading the bulb.


2. The position or size of any hole on the top cover of the medical device shell shall ensure that the test rod with a diameter of 4 mm and a length of 100 mm will not touch live parts when it is freely and vertically suspended in the access hole throughout its length.


3. When the handle, knob, control lever, etc. are removed, the conductor part of the operating mechanism of the control device can be touched. When the open circuit test voltage is not greater than 50V AC and the test current is not less than 1a, the resistance measured to the protective grounding terminal should not be greater than 0.2 Ω; Or it should be isolated from live parts by adopting a method of 17g in GB 9706.1-2007.


4. If the components with line voltage above 25V AC or 60V DC in the shell of the medical device cannot be disconnected from the power supply through the external main switch or the plug device that can be pulled out at any time, additional covers shall be used for protection, so that contact can be prevented even after the shell is opened, or when the spaces are separated and arranged, the "live" mark shall be clearly marked.


5. In order to prevent contact with live parts, the shell must be removed only with tools, otherwise an automatic device should be used when opening or removing the shell, so that these parts are not used.


6. When the device has a preset controller, it can be adjusted by the user with tools during normal use. The hole opened for adjusting these control devices shall be designed so that the adjusting tool cannot contact the basic insulation or any live parts in the hole, or any parts isolated from the grid power supply part only through the basic insulation but without grounding protection.


Copyright: Shenzhen Zhongyuan Plastic Co., Ltd. | mold processing customized manufacturer. If you want to know the processing price of the factory, how much is it, and which is good, please contact us Yue ICP Bei No. 17060480
主站蜘蛛池模板: 欧美一级一级做性视频 | 亚洲一区二区三区四区视频 | 亚洲欧美视频在线观看 | 亚洲 欧美 在线观看 | 欧美精品久久久久久久久大尺度 | 欧美性猛交xxx免费看人妖 | 久久亚洲国产伦理 | 国产亚洲免费观看 | 亚洲免费成人网 | 亚洲一区第一页 | 欧美色综合天天综合高清网 | 国产日韩欧美第一页 | jjzz日本护士 | 日本免费小视频 | 久久天天躁狠狠躁夜夜avai | 久碰人澡人澡人澡人澡91 | 99免费在线 | 香港台湾日本三级纶理在线视 | 五月天三级 | 欧美午夜小视频 | 偷拍肉窝窝视频在线播放 | 精品久久久久久久高清 | 色婷婷天天综合在线 | 日韩免费高清一级毛片在线 | 四虎网站网址 | 欧美在线观看不卡 | 欧美日韩综合精品一区二区三区 | 久久综合第一页 | 成人观看网站a | 91精品国产自产在线观看 | 国产亚洲第一精品社区麻豆 | 女bbbbxxxx毛片视频丶 | 日本高清无卡码一区二区久久 | 四虎综合网 | 国产专区路线 | 女人被狂躁的视频免费动图 | 久久成人国产精品青青 | 日韩美一区二区 | 国产一级黄色网 | 日本免费观看95视频网站 | 亚洲综合色播 |